CM2 |
Set Maximum2 |
CM2 |
Command Group: Calibration | Access: | Network interfaces: | ||
| TAC protected | RW | CANopen | SDO2300sub0E | |
| MODbus | 221A,2 | |||
CM2 - Set the maximum allowable output value #2. | PROFIbus | read 3,14 write 4,14 | ||
|
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
CM2 - Maximum display value for range 2.
If CM2 = 0 then CM2 is disabled.
Set the maximum allowable output value in interval or range 2.
Lower limit CM1+1, upper limit 999999.
Issuing the command without any parameters returns the current CM2 value.
This value will determine limit for range 2. When the weight becomes greater than CM2 and CM2<CM3 then range 3 is entered, else the output will change to ooooooo,
signifying over-range.
To set a new value for CM2,
the command must be preceded by the CE XXXXX command, where XXXXX is the current TAC.
The new CM value required is then input as a parameter of CM, in the format CM 2 4010.
For further information, please see Calibration Procedure.
Factory default setting: Max 1 = 999999, Max 2 = 0, Max 3 = 0.
CM2 is implemented in LDU78 from Firmware 78.183.v.2.47
See also CE, CI, DS, DP, CZ, CG, ZT, AZ, CS, FD, UR, MR, CM1, CM3.