LN

Lin point

LN


Command Group: Calibration

Access:

Network interfaces:

TAC protected

RW

CANopen

MODbus

LNn - Set Liniarization points.

PROFIbus


Synopsis:
CommandResponse
GetLNL<Value>
SetLN <Value>OK (or ERR)
Valid forMin. valueMax. value
151.181    
152.181    
69.181    



LNn Linearize – this command make it possible to inspect and correct more than the two normal calibration points set by the CZ and CG command – up to seven calibration points are allowed. Issuing the command with only one parameter (1-7) returns the calibration point in the format: Ln:XXXXXXX YYYYYYY where XXXXXXX is the input level in mV/V (without decimal point) and YYYYYYY is the displayed number of increments. Calibration can be done in the following way: LN1 0 0 define zero output at zero mV/V input. LN2 200000 100000 defines 100000 increments at +2mV/V input – at least two points must be given.
Changes in the linearization table require a preceding CE command.
NOTE: Before a recalibration with fewer points than the previous one a LC command must be issued.

See also LC, CE, CZ, CG.