LE

Get Last Error

LE


Command Group: Standard

Access:

Network interfaces:

R

CANopen

SDO1002sub0

MODbus

3004,2

LE - Get Last Error

PROFIbus


Synopsis:
CommandResponse
GetLEE:<Value>
Valid forMin. valueMax. value
141.181    
141.184    
141.185    
141.186    
143.181    
143.184    
143.185    



LE Last Error.

Returns the most recent error occurred.
The values are:

CodeNameText
000NO ERROR 
001NOT_IMPLEMENTEDThe command is unknown to the device.
002NOT_READYThe device cannot execute the command now.
003ERR_BAUDThe device does not support the requested baud rate.
004CAL_NOT_OPENThe command cannot be executed because the TAC protection is on.
005BAD_CAL_IDThe requested Calibration parameter Field is not present in the device.
006BAD_CAL_VALUEThe value for the Calibration parameter Field is outside the allowed range.
007TIMEOUTThe command was terninated because the weight was unstable for too long.
008NOT_STABLEThe command was rejected because the weight was not stable.
009BAD_FILL_PARAM_IDThe requested Filling parameter Field is not present in the device.
010BAD_FILL_PARAM_VALUEThe value for the Filling parameter Field is outside the allowed range.
011BAD_GEN_VALUE_IDThe requested General parameter Field is not present in the device.
012BAD_GEN_PARAM_VALUEThe value for the General parameter Field is outside the allowed range.
013BAD_TRIG_VALUE_IDThe requested Trigger parameter Field is not present in the device.
014BAD_TRIG_PARAM_VALUEThe value for the Trigger parameter Field is outside the allowed range.
015BAD_TARE_RANGEThe requested Tare value is outside the allowed range.
016BAD_FILL_SLOPEThe filling slope can neither be determined nor monitored.
017BAD_FLOW_VALUE_IDThe requested Loss-In-Weight/Flow parameter Field is not present in the device.
018BAD_FLOW_PARAM_VALUEThe value for the Loss-In-Weight/Flow parameter Field is outside the allowed range.
019ZEROING_DISABLEDThe Zeroing function has been disabled in this device.
020OUT_OF_ZERO_RANGEZeroing this weight is not permitted.
021NOT_ENOUGH_RESOLUTION 
022INPUT_RANGE_EXCEEDEDThe electrical weight input is outside the specifications for the device.
023LOAD_CELL_CONNECTION_ERRORThe loadcell electrical inputs is not making sense.
024COMMAND_NOT_ALLOWEDIf you try to E.g. "Set Zero" while device is doing an average measurement.